FBXXX Field Temperature Block
Non-SCPI Commands by Function or Group
68
s [n]
Read or Set the temperature control set-point in °C or °F (based on current system
units). Where “n” is a real value with acceptance limits based on the model. If “n” is
left blank, the command will be treated as a query. This query responds with the com-
mand string “set:” followed by the temperature setting and associated units. Refer to
SOUR:SPO command.
Read Example: s
set: 100.00 C
Set Example: s 250
sa [n]
Read or Set the serial interface auto printing interval. Where “n” is an integer value
from 0-60. If “n” is 0, the auto print will be disabled. Values range from 1 to 60 and
are in seconds. If “n” is left blank, the command will be treated as a query. This query
responds with the command string “sa” and a Colon followed by the interval setting.
Read Example: sa
sa: 5
Set Example: s 10
t
Read the control temperature in °C or °F (based on current system units). This com-
mand is query only and responds with the command string and a Colon followed by
the temperature and associated units. Refer to SOUR:SENS:DAT command.
Read Example: t
t: 99.988 C
u[n]
Read or Set the display temperature units, where “n” is a character “C” of “F”.
Default: C If “n” is left blank, the command will be treated as a query. This query
responds with the command string “u” and a Colon followed by the unit setting. Refer
to UNIT:TEMP command.
Read Example: u
u: C
Set Example: u F