EasyManua.ls Logo

BEP 600-DCSM - Configuration; DC Voltage Configuration

BEP 600-DCSM
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Configuration Page 15
3. Configuration
DC Voltage
Calibration
Voltage Calibration is factory set. If you need to alter the calibration follow the steps below:
Go to Main Menu > Setup > Input(s) > ‘Choose Voltage Input’ > Calibration.
Step 1
Insert a ground wire into the pin of the input you are configuring. Select ‘Zero’ on the meter and
then select ‘Yes’.
Step 2
Remove ground and then wire the Input (DC Positive) into the same pin.
Step 3
Check battery voltage using a calibrated voltmeter then select ‘Actual’ and enter the correct voltage
and then select ‘Save’.
Settings
Go to Main Menu > Setup > Input(s) > ‘Choose Voltage Input’ > Settings
DC Voltage Value
Label