10 Bertec Corporation Instrumented Treadmill Manual
Proper power should be available at the installation site so that the treadmill can be tested after installation.
Power Requirements
Bertec Instrumented Treadmills need standard industrial 3-phase power supply to operate.
Three components of the treadmill system must be supplied with power: the treadmill control cabinet and both ampliers. The
cabinet will connect to its power supply using a straight bladed plug specied by Bertec. The cabinet should be within easy reach
of the power source. Each amplier takes power from a standard wall outlet, 120 V in North America for example. The treadmill
is operated using one control input. This is a B-type USB connector located on the AM6500-TM unit.
The voltage and current requirements change depending in which country the customer is located. The customer is responsible
for supplying the correct power with the proper wall receptacle for the treadmill. The power plugs and main power supply
necessary to run the standard speed dual treadmill is given in the following table:
FIGURE 3: Instrumented treadmill power requirements
Location Voltage/Phase (V) Current Phase (A) Frequency Plug on Treadmill
North America
208±10%
40 A 60 Hz 15-50P*
Japan
200±10%
40 A 50/60 Hz 15-50P*
Europe
400±10%
20 A 50 Hz 532P6W**
Australia
400±10%
20 A 50 Hz 56PA532
*NEMA Standard
**IEC-60309 Standard
For power requirements in other locations, please consult Bertec Corporation.