FDL (E2.1) 05/2016 page 57/75
16. Troubleshooting
Set-point temperature is not
reached after specified time.
LED “AIR” (3) is not lit.
Door gasket defective. Replace door gasket,
Controller not adjusted. Calibrate and adjust controller.
Chamber heating permanently,
set-point not maintained.
Contact BINDER service.
Semiconductor relay defective
Calibrate and adjust controller.
Chamber doesn’t heat up.
LED “AIR” (3) is not lit.
LED (2a) of safety device is lit.
Controller display off.
Fan turning.
Safety device has turned off
the oven.
Limit temperature reached.
Safety device class 2 (chap.
12.1) set too low.
Allow the oven to cool down the
oven and press the “RESET” button.
Check temperature set-point and
setting of safety device (chap. 12.1).
If appropriate, select suitable limit
Semiconductor relay defective.
Contact BINDER service.
Chamber without any function.
No power supply.
Check connection to power supply.
Check if the main power switch (1)
is turned on.
Deviations from the indicated
Oven fully loaded.
Charge the oven less or consider
Chamber doesn’t heat up.
LED “AIR” (3) is lit.
Indication “RESET ALARM” in
Display 2.
Acoustical signal (can be reset on
the controller)
With pushbutton “START” (4)
the prepurging time has been
started. No heating release
Wait approx. 2 minutes.
Chamber door not closed.
Close the chamber door completely,
press pushbutton “START” (4) and
Condition following a power
failure.
Press pushbutton “START” (4) and
wait approx. 2 minutes.
Exhaust air channel blocked.
Check exhaust air system (customer
side).
Intake opening blocked
Change or clean filter.
No volumetric flow rate for fresh
air and forced-air circulation.
LED “AIR” (3) is lit.
Indication “RESET ALARM” in
Display 2.
Acoustical signal (can be reset on
Chamber door not closed.
Close the chamber door completely,
press pushbutton “START” (4) and
Condition following a power
failure.
Press pushbutton “START” (4) and
wait approx. 2 minutes.
Fan defective. Contact BINDER service.