Binder FDL 115 - 17.3 Technical Data

Binder FDL 115
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FDL (E2.1) 05/2016 page 60/75
Interior dimensions
Width
mm / inch
600 / 23.62
Height
mm / inch
435 / 17.13
Depth
mm / inch
435 / 17.13
Interior volume
l / cu.ft
115 / 4.1
Total steam space volume
l / cu.ft.
156 / 5.51
Racks
Quantity of racks (regular)
2
Quantity of racks (max.)
5
Maximum load per rack
Kg /lbs
30 / 66
Maximum permitted total load
Kg /lbs
60 / 132
Weight
Weight of the chamber (empty)
Kg /lbs
88 / 194
Temperature data
Temperature range, 10 °C above ambient up to
°C / °F
300 / 572
Maximum temperature of the heating surface
°C / °F
750 / 1382
Maximum permitted oven temperature
°C / °F
300 / 572
Temperature fluctuation
at 150 °C / 302 °F
± K
0.8
Temperature uniformity (variation)
at 50 °C / 122 °F
± K
0.5
at 150 °C / 302 °F
± K
2.5
at 300 °C / 572 °F
± K
5.8
Heating-up time
to 50 °C / 122 °F
min
15
to 150 °C / 302 °F
min
20
to 300 °C / 572 °F
min
60
Recovery time after
door was opened
for 30 sec.
to 50 °C / 122 °F
min
8
to 150 °C / 302 °F
min
12
to 300 °C / 572 °F
min
20
Ventilation data
Air circulation
approx. x/min.
20
Air change acc. to EN 1539:2015 at 50 °C
approx. x/min.
2.5
Volumetric flow rate of exhaust air acc. to EN 1539:2015
at 50 °C
approx. l/min
400
approx. m
3
/h
24.0
Solvent data
Highest permitted solvent quantity acc. to EN 1539:2015
(at T-180 °C, M-100g/mol, U-40g/m3, K=0,5)
g 6.65
Electrical data
Housing protection acc. to EN 60529
IP
33
Nominal voltage (+5 %)
at 50 Hz power frequency
V
230
at 60 Hz power frequency
V
230
Current type
1N~
Nominal current
Amp
13.0
Nominal power
kW
2.90
Power plug
shock proof plug
Over-voltage category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Environment-specific data
Noise level (mean value)
dB(A)
57
Energy consumption
at 150 °C / 302°F
Wh/h
1098