c-6. Type setting
❖ With the keys or , select "Type", the current value is displayed on the side (by default: PROP)
❖ Press again: the value flashes and can be modified
❖ With the keys or , set the desired new value and then press to validate
d.) Flow
As the apparatus must control the filtration through the circulating pump control cable, the Flow setting
must always be set to ON.
e.) Calibration mode
• With the keys or , select "Cal", the current value is displayed on the side (by default: Full)
• Press again: the value flashes and can be modified
• With the keys or , set the desired new value and then press to validate
➢ RECOMMANDED : Full (Buffer solutions kit: pH 4, pH 7, Redox 465 mV)
➢ Easy (Buffer solutions kit: H 7, Redox 465 mV)
➢ Off (Disabled)
• Press to save the value and return to “Program Menu”
f.) Password
• With the keys or , select "Password", **** is displayed on the side
• Press : the current value displays (by default: 0000). If this value remains at 0000, the password is
disabled. Any other value will activate the parameters protection by a password
• Press to modify the digit value and press to move to another digit
• Press to save the value and return to “Program Menu”
5. Saving the setting(s)
• From the “Program Menu”, press
• With the keys or , select « Save » then to save and exit