K. REDOX CALIBRATION PROBE
1
2
3
4
Wash the probe
Hold the probe in the buffer
solution
5
6
7
8
Press until the
message "CALIBRATION"
is displayed
The message
"Press_CAL"
"Buffer sol 465mV"
is displayed
Press
The message
"Wait 60 s"
"Buffer sol 465mV"
is displayed
If the message
« 465mV Quality’ 100% »
is displayed, go to step
8.
If the message
"Change the probe"
is displayed, press
to exit.
Change the probe and
return to step 1.
Wash the probe
9
10
11
Press to exit and
save the data
Displays the redox
measure in the pool