EasyManua.ls Logo

Biobase BSC-4FA2 - Technical Specifications and Performance

Biobase BSC-4FA2
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
2.1.4 Protected objects
Microbiological Safety Cabinets (BSCs) are designed to protect the operator, the laboratory
environment and work materials from exposure to infectious aerosols and splashes that may be
generated when manipulating materials containing infectious agents, such as primary cultures, stocks
and diagnostic specimens.
2.1.5 TECHNICAL PARAMETERS
Model
Parameters
BSC-4FA2(4’)
Power Supply
220V 50/60HZ
External Size(W*D*H)
1383×775×2295mm
Working Zone
Size(W*D*H)
1210×600×660mm
Total Airflow Volume
465m3/h
Downflow Velocity
0.33m/s
Inflow Velocity
0.53m/s
Fluorescent lamp
Consumption
28W*2
UV Lamp Consumption
30W
HEPA Filter
99.999%(Diameter:0.3μm
Noise
≤65dB
A
Notes :(1) Electric consumption power including two powers which operation area needs to load,The
power of a single load is 230V*1.5A
(2) The Company reserves the right to change the design of product, we will change design of
product without prior notice.
2.1.6 Performance Index
1) Biological safety functions
Personnel protection, microbial colony count ≤5CFU;
Sample protection, microbial colony count ≤5CFU;
Cross contamination protection, microbial colony count ≤2CFU.