door shouldn’t be against over door or window,but far away from Ventilation and air conditioning
vents.In order to against outside various high-speed current of air to interference of operation area,
such as door, window, fan, draught of sending air, compressed air etc.
Test shows that if other interference airflow exceeds the intake airflow rate of the safety ca binet,
the indoor air will enter the working area of the cabinet.So it is very necessary to place the cabi net in
the correct position.It should also pay attention to the relationship between the safety cabinet e
xhaust air and indoor ventilation airflow or exhaust duct.Exhaust air discharged from the top , When
placing the cabinet, it should avoid restricting its exhaust.Biological safety cabinets should be in the
direction of the air flow downstream and the side of the security cabinet at least 300mm space to help
check.
Working environment:
(1) Only is suitable for indoor;
(2) Altitude not exceeding 2000m;
(3) Ambient temperature: 15
℃~
35
℃
;
(4) Relative Humidity: ≤75%;
(5) Atmospheric pressure range: 70 kPa~106 kPa;
(6) Electrical parameters: Consistent with the rated voltage of the Microbiological Safety Cabinet
(See 2.1.5 technical parameter performance index);
(7) Power supply need to be grounded; (Judging method: testing the fire wire and the zero line of the
power supply with multimeter, the fire wire to ground voltage should be grid voltage and the zero
line to ground voltage should be 0, otherwise the power supply ground is bad).
1.4 Installation
a. Remove all the package materials;
b. Inspect the surface of main body to make sure whether there is scratch, deformation or
uncorrelated things;
c. Chor forklift;eck the attachment and information accroding to the packaging list in insruction.
d. Move the whole device to the final installation location,where is convenient to install.
e. This equipment shouldn’t be hosting.Wooden packaging should be used as tap support in
transpor process, carried by hydraulic truck or forklift.
f. The whole power line 3 * 1.0mm² * 3000mm; the whole ground line is yellow-green 14AWG
The base stand will be packed at back of main body, please take it out before installation.
DO NOT INVERT, DISASSEMBLE OR TITILE THE CABINET duringtransportation.