Before fumigation should’t open the cabinet enclosure, prohibit removing any partsr
from the Microbiological Safety Cabinet!
2.6.2 Clean the cabinet surface
1) Clean the operating area surface
Wipe the entire surface with a soft cotton cloth or towel soaked with concentrated liquid soap, then
wipe up the soap with another cotton cloth or towel soaked with clean hot or warm water, and then
wipe the surface with a dry cotton cloth or towel rapidly.
Cleaning method: When the test substance is accidentally spilled on the work surface, firstly with a
paper towel to wipe up, then wash and clean with water or other materials (Notice:! Cleaning
material used does not set off a chemical reaction with the test substance) , waste liquid discharges
from the bottom drain valve in the front left . After wiping twice, then use the solution of the test
lytic substance (Notice:! The solution can used doesn’t set off a chemical reaction with the test
substance) to wash work surface, while using a hair brush cleaning twice, then wipe and wash with
water or other materials , waste liquid discharges from the sump drain valve in the left front, then
clean the sump.
Before fumigation should not open the cabinet enclosure, prohibit removing any
partsr from the Microbiological Safety Cabinet!
Fumigation method : Safety cabinet should be sealed prior before fumigation to ensure that no
formaldehyde leak into the laboratory or another room. Fumigation for safety cabinet and filter
should ensure that export of HEPA filter and piping systems exposed for more than 14 hours in order
to ensure that microorganisms that pierce the filter is inactivate . When half of formaldehyde is
evaporated, turn on cabinet fan running for 10s ~ 15s, making the whole area filled with
formaldehyde; After complete evaporation , and then turn on the cabinet fan once more running
for 10s ~ 15s. After the fumigation, Microbiological Safety Cabinet sealing device should be
removed and let the exhaust fan on for not less than 8 hours, so before using or servicing the cabinet
to remove residual formaldehyde vapor.
There are signs of contamination, work surface,sump and etc,we can use medical alcohol or other