contact service personal.
(11
Front Grille is used for air intake and drain. Do not block it, otherwise it will affect airflow.
Armrest is recommended to solve this problem and reducing the operator's wrist fatigue.
(12
Long-term use of Microbiological Safety Cabinets will inevitably cause pollution (e.g. HEPA
filters, corner cabinets, etc.). In order to sterilize thoroughly every 500 hours, formalin
(formaldehyde) fumigation sterilizer is recommended. After sterilization, neutralize
formaldehyde gas with ammonium hydrogen carbonate. Make sure no sterilization gas
escapes during the whole process.
(13The maximum storage period is one year. If the period is more than one year, performance
test should be done.
(14
In transportation process should take appropriate protective measures according to diagram
on the surface of package. Purchaser in receiving should carefully check the integrity of the
box .if there is damage and crush, please refuse to accrpt. And timely contact with our
company.
(15When Microbiological Safety Cabinet installed and used move again, please timely contact
with our company.
(16The ground bearing requirements: ground bearing capacity ≥450KPa.
(17Total weight: 220Kg.
(18
Allowable pressure of water socket: ≤0.8MPa. (Optional)
(19For experiments that produce waste liquid, please timely drain waste water in the sump
through drain valve. If the liquid waste is harmful to human’s body, according to the grade of
pollution should add the corresponding grade container, or connect to the corresponding
grade drain pipes, and drain waste into the experimental equipment to immediately dispose.
(20Person can operate Microbiological Safety Cabinet only if he is qualified after training,
during using Microbiological Safety Cabinet if it is a harmful experiment, operator must wear
appropriate protective gloves, masks and lab coats, etc., and avoid touching mouth , eyes and
face.
(21
There is risk to move the parts moving .Relevant personnel can use this equipment must after
training. while the front window must be raised to 200mm and in this case person can operate
in the cabinet. In this way we can reduce harm out of front window trouble.
(22
When the equipment is repaired or scrapped. The internal parts can’t be thrown about because
of biological hazards .please deal with them according to local regulations.
Serious declaration: we will take no responsibility for risks caused by improper operation