2)Replace the UV lamp
UV lamp should be replaced regularly according to the frequency of use, when using UV lamps
reach to the time of 600 hours, we recommend to replace the lamp. When replacing, first make sure
the power is off, and then screw the bulb 90 °and take it off , then take the correspondence type of
lamp, and put it to the lamp holder and and screw 90 °in reverse direction. After replacing the UV
lamp, it needs to keep pressing the button of UV for about five seconds when the machine stays
standby.
Picture 16
2) Replace the fuse
Power plug fuse is located in the top operation panel,When replace it, turn off the power and
disconnect plug,and take out of the fuse holder using a slotted screwdriver and replace with
a new fuse, and then press it back
Picture 17