Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 48
Screen
Cycle Name Entry
The name of the cycle may be changed.
Function keys:
<F1> or <F2>
Select "Chg" to change the name for the cycle.
Screen
Cycle Name Entry
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"Sel" to select the current letter/number.
<+> or <->
Use these keys to scroll through the available letters (A-Z) and
numbers (0-9)
<enter>
Confirm the name of the cycle and go to the next screen.
Screen
Aspiration Time Entry
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"<<" to return to the previous screen.
<+> or <->
Use these keys to change the aspiration time (the time the manifold
remains inside a well during aspiration) in steps of 1sec (range: 0 - 10
seconds).
<enter>
Confirm the selection of the aspiration time and go to the next screen.
Aspiration time:
<< Esc 5 sec