Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 49
Screen
Aspiration Power Entry
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"<<" to return to the previous screen.
<+> or <->
Use these keys to change the power with which the liquid is aspirated
off a well (“Low”, “Medium”, “High”).
<enter>
Confirm the selection of the aspiration power and go to the next
screen.
Screen
Dispense Volume Entry
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"<<" to return to the previous screen.
<+> or <->
Use these keys to change the dispense volume for this cycle in steps of
50µl (range 0 - 2000µl).
<enter>
Confirm the selection of the volume and go to the next screen.
Screen
Dispenser Power Selection
Disp volume:
<< Esc 1000 µl