Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 58
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes or
"<<" to return to the previous screen.
<+> or <->
Use these keys to toggle between "Yes" and "No".
<enter>
Confirm the selection for crosswise aspiration and go to the next
screen.
Screen
Side Position Front Entry
In this menu the side position front for a flat-bottom plate is defined. If
you choose crosswise aspiration this is the first aspiration position. Else
this is the aspiration position: for a better aspiration result it is
recommended to position the needle close to one wall. The side position
front may be defined by "teach-in" (move the plate to the correct position
with the <+> and <-> keys).
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes.
Select “Mov” to position the plate to the current side position front and
switch to the “teach-in” mode.<+> or <->
Use these keys to enter the side position front. The plate can be moved
in steps of 0.1mm (range: -5.0 – 5.0mm).
<enter>
Confirm the side position front and go to the next screen.
Screen
Side Position Front “Teach-in” mode
In this "teach-in" you can move the plate to the correct position with the
<+> and <-> keys.
Function keys:
<F1> or <F2>
Select "" to automatically detect the bottom of the well (changes to
Side pos front:
Esc Mov 0.1 mm
Side pos front:
Mov 0.1 mm