Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 59
“”).
Select "" to move the manifold up.
Select “Mov” to move the plate transport to the current side position
front.
<+> or <->
Use these keys to enter the side position front. The plate can be moved
in steps of 0.1mm (range: -5.0 – 5.0mm).
<enter>
Confirm the side position front and go to the next screen.
Screen
Side Position Rear Entry
In this menu the side position rear for a flat-bottom plate is defined. It is
only available if crosswise aspiration is enabled and it defines the second
aspiration position. The side position rear may be defined by "teach-in"
(move the plate to the correct position with the <+> and <-> keys).
Function keys:
<F1> or <F2>
Select "Esc" to return to the main menu without saving any changes.
Select “Mov” to position the plate to the current side position rear and
switch to the “teach-in” mode.
<+> or <->
Use these keys to enter the side position rear. The plate can be moved
in steps of 0.1mm (range: -5.0 – 5.0mm).
<enter>
Confirm the side position rear and go to the next screen.
Screen
Side Position Rear “Teach-in” mode
In this "teach-in" mode the plate may be moved to the correct position
with the <+> and <-> keys.
Function keys:
<F1> or <F2>
Select "" to automatically detect the bottom of the well (changes to
“”).
Side pos rear:
Esc Mov 0.1 mm
Side pos rear:
Mov 0.1 mm