Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 67
Screen
Waste Setup
Use the <-> / <+> keys to select whether a liquid level sensor is attached
to the waste bottle and confirm with <enter>. To use this function a liquid
level detector must be connected to the instrument and the waste bottle.
If activated the liquid level of the waste bottle will be checked before the
procedure is started. In case that the level of the liquid in the waste bottle
is above the sensor a corresponding warning will be displayed.
Screen
Rinse Setup
Use the <-> / <+> keys to select whether a liquid level sensor is attached
to the rinse bottle and confirm with <enter>. To use this function a liquid
level detector must be connected to the instrument and the rinse bottle.
If activated the liquid level of the rinse bottle for a started procedure will
be checked before the procedure is started. In case that the level of the
liquid in the rinse bottle is below the sensor a corresponding warning will
be displayed.
Screen
Wash1 Setup
Use the <-> / <+> keys to select whether a liquid level sensor is attached
to the wash1 bottle and confirm with <enter>. To use this function a liquid
level detector must be connected to the instrument and the wash1 bottle.
If activated the liquid level of the wash1 bottle for a started procedure will
be checked before the procedure is started. In case that the level of the
liquid in the wash1 bottle is below the sensor a corresponding warning
will be displayed.
Waste full det:
<< Esc Off
Rinse empty det:
<< Esc Off
Wash1 empty det:
<< Esc Off
Wash2 empty det:
<< Esc Off