Biochrom Anthos Fluido-2 Microplate Washer v2.0 Page 68
Screen
Wash2 Setup
Use the <-> / <+> keys to select whether a liquid level sensor is attached
to the wash2 bottle and confirm with <enter>. To use this function a liquid
level detector must be connected to the instrument and the wash2 bottle.
If activated the liquid level of the wash2 bottle for a started procedure will
be checked before the procedure is started. In case that the level of the
liquid in the wash2 bottle is below the sensor a corresponding warning
will be displayed.
Screen
Wash3 Setup
Use the <-> / <+> keys to select whether a liquid level sensor is attached
to the wash3 bottle and confirm with <enter>. To use this function a liquid
level detector must be connected to the instrument and the wash3 bottle.
If activated the liquid level of the wash3 bottle for a started procedure will
be checked before the procedure is started. In case that the level of the
liquid in the wash3 bottle is below the sensor a corresponding warning
will be displayed.
Screen
Liquid Warning Setup
Use the <-> / <+> keys to activate/deactivate the liquid warning and
confirm with <enter>. If activated a reminder to check the level of the
wash bottle(s) is displayed prior to the start of a procedure.
Screen
Waste Warning Setup
Wash3 empty det:
<< Esc Off
Liquid warning:
<< Esc On
Waste warning:
<< Esc On Yes