E80/E70 Digital Electrocardiograph User’s Manual
2-2
risk of entanglement or strangulation by patient or personnel.
Keep the packing materials out of children’s reach. Disposal of the
packing materials should observe the applicable waste control
regulations.
Although safe request is considered during device design, but
operator must not ignore device status and observing patient. Please
especially notice device and patient can’t be moved during device
working.
Device connected to digital and stimulant connector must be validated
with each IEC standards (e.g. data processing equipment standard:
IEC950, Medical electrical equipment: IEC60601-1), and all the
configurations must comply with availability version of IEC60601-1.
so the medical system must comply with availability version of
IEC60601-1.
The electrocardiograph can be directly applied to heart.
Caution:
At the end of its service life, the cardiograph, as well as its accessories,
must be disposed of in compliance with the guidelines regulating the
disposal of such products. If you have any questions concerning
disposal of the cardiograph, please contact us.
Magnetic and electrical fields are capable of interfering with the
proper performance of the cardiograph. For this reason make sure
that all external devices operated in the vicinity of the cardiograph
comply with the relevant EMC requirements. Mobile phone, X-ray
equipment or MRI devices are a possible source of interference as
they may emit higher levels of electromagnetic radiation.
Before connecting the cardiograph to the power cord, check that the
voltage and frequency ratings of the power cord are the same as those
indicated on the cardiograph’s label or in this manual.
Always install or carry the cardiograph properly to avoid damage
caused by drop, impact, strong vibration or other mechanical force.