E80/E70 Digital Electrocardiograph Users Manual
2-3
Note:
Put the cardiograph in a location where you can easily see the screen
and access the operating controls.
Keep this manual in the vicinity of the cardiograph so that it can be
obtained conveniently when needed.
The software was developed in compliance with IEC 60601-1-4. The
possibility of hazards arising from software errors is minimized.
This manual describes all features and options. Your cardiograph may
not have all of them.
2.2 General Safety
Warning: The cardiograph is neither a therapeutic instrument nor a
device that can be used at home.
1Safety precautions for installation
Connect the power cord to a properly grounding socket. Avoid putting the
socket used for it in the same loop of such devices as the air conditioners,
which regularly switch between on and off.
Avoid putting the cardiograph in the locations where it easily shakes or
wobbles.
Enough space shall be left around the cardiograph so as to guarantee
normal ventilation.
Make sure the ambient temperature and humidity are stable and avoid the
occurrence of condensation in the operation process of the cardiograph.
Warning: Never install the cardiograph in an environment where
flammable anesthetic gas is present.
2Cardiograph conforms to the safety requirements of IEC 60601-1:1988+
A1:1991+A2:1995. This cardiograph is protected against defibrillation
effects.
3Notes on symbols related to safety
Type CF applied part, defibrillation protected