E80/E70 Digital Electrocardiograph Users Manual
7-3
2 Chest electrode and limb electrode
a) Please clean chest electrode and limb electrode before disinfection.
b) Wipe off the conductive paste on surface of electrode with soft cloth;
c) Take apart the electrode plate and clamp of limb electrode as well as
rubber ball and metal cup of chest electrode;
d) Put electrode into clean warm water (not higher than 35) and clean
it to ensure no residue of conductive paste;
e) Air dries the electrode at a shady and cool place.
Caution:
Electrode shall be timely cleaned after use;
Rubber ball of chest electrode shall be prevented from direct sunlight;
otherwise, aging will be caused;
Electrode with eroded surface shall be timely replaced with new
electrode.
Caution:
Do not disinfect cable and lead wire with high voltage, radial or steam.
Do not dip cable or lead wire directly in liquid.
To prevent long-term damage to cable, it is suggested that the product be
disinfected only when your hospital regulation deems it as necessary.
Do not clean or reuse disposable electrode.
Warning: Do not use EtO, phenyl, amido or iodo for disinfection of
the machine.
Caution: Disinfection possibly causes damage to Electrocardiograph
to some extent. It is suggested that disinfection be conducted only
when your hospital maintenance plan deems it as necessary. The
equipment shall be cleaned before disinfection.
Note: Electric schematic diagram and list of components are provided to
the qualified maintenance station or personnel certified by the manufacturer
only.