BM3 www.BIO2NET.com
Revision A
- 12 -
Pulse Width: 100 ±10 ms
Delay from R-wave peak to start of pulse: 35 ms, maximum
Short circuit current: 15 mA Minimum required R-wave
amplitude: 0.5 V
– LAN digital output for transferring data to an external computer
– nurse call
• Battery (standard)
– internal battery: sealed lead-acid
– battery status indicator
– Operating Time: 2.5 hours typical (fully charged battery)
• Thermal Printer (Optional):
– Speeds: 25, 50 mm/sec
– Paper Width: 58 mm
Graphical and Tabular Trends
• Tabular Trends
– Memory Storage: 24 hours
– Data Interval: 1 minute
– Display Interval: 1MIN, 5, 15, 30, 1HR
Tabular Format: One table for all variables
• Graphical Trends
- Display Period: 30MINS, 60, 90, 3HRS, 6, 12
SpO
2
Performance Specifications
• % Saturation Range: 0% to 100%
• Pulse Rate Range: 30 to 300 bpm
• SpO2 Accuracy: 70% to 100% ± 2 digits, 0% to 69% unspecified
• Pulse Rate Accuracy: ± 3 bpm
NIBP Performance Specifications
• Technique: oscillometric
• Measurement Modes:
– Manual:Single measurement
- Auto:Automatic intervals of 1MIN,2,3,4,5,10,15,20,30,1HR,2,4,8
• Cuff Pressure Display: 30 to 300 mmHg
• Blood Pressure Measurement Range: