BM3 www.BIO2NET.com
Revision A
- 13 -
– Systolic: 60 to 250 mmHg
– Mean Arterial Pressure: 45 to 235 mmHg
– Diastolic: 40 to 220 mmHg
• Pulse Rate Range: 40 to 200 bpm
ECG Performance Specifications
• Leads: 3 leads
• Heart Rate Range: 30 to 300 bpm
• Heart Rate Accuracy: ± 3 bpm
• Bandwidth: 0.5 Hz to 40 Hz
• Display Sweep Speeds: 25mm/sec
• ECG Size (Sensitivity): 0.5, 1, 2, 4 mV/cm
• Lead-Off Detection with display indicator
• Pacemaker Detection Mode: Indicator on waveform display,
user selectable
• Differential Input Impedance: > 5 Mohm
• Common Mode Rejection Ratio: > 90 dB at 50 or 60 Hz
• Input Dynamic Range: ± 5 mVAC, ± 300 mVDC
• Defibrillator Discharge: <5 s
• Defibrillation Artifact Recovery Time: < 8 s
Respirations Performance Specifications
• Range: 5 to 120 breaths/min
• Accuracy: ± 3 breaths/min
• Display Sweep Speeds: 25mm/sec
Temperature Performance Specifications
• Range: 15°C to 45°C (59°F to 113°F)
• Accuracy: 25°C to 45°C ± 0.2°C, 15 °C to 24°C ± 0.3°C
• Compatible with YSI Series 400 Temperature Probes
Accessories included:
- 3-lead patient cable 1ea
- electrodes 10ea