EasyManua.ls Logo

BMG Labtech FLUOstar Omega - Technical Specifications; Measurement Principles; Reagent Injection

BMG Labtech FLUOstar Omega
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BMG LABTECH Omega Operating Manual
2014-05-16 0415B0001G 5/29
1 Technical Specifications
Measurement principles
FLUOstar Omega: - Fluorescence intensity
- Time-resolved fluorescence TR-FRET (opt.)
- Luminescence (opt.)
- UV/VIS absorbance spectra
- AlphaScreen
®
/ AlphaLISA
®
(opt.)
- Upgradeable to POLARstar Omega
LUMIstar Omega: - Luminescence
- Simultaneous dual emission luminescence (opt.)
- Upgradeable to FLUOstar / POLARstar Omega
POLARstar Omega: - Fluorescence polarization
- Fluorescence intensity
- Simultaneous dual emission fluorescence
- Time-resolved fluorescence TR-FRET (opt.)
- Luminescence (opt.)
- Simultaneous dual emission luminescence (opt.)
- UV/VIS absorbance spectra
- AlphaScreen
®
/ AlphaLISA
®
(opt.)
SPECTROstar Omega: - UV/VIS absorbance spectra
- Upgradeable to FLUOstar / POLARstar Omega
Light source
FLUOstar, POLARstar & SPECTROstar Omega: High-energy xenon flash-lamp
Detector - Side window photomultiplier tube
Filters
FLUOstar & POLARstar Omega: - 2 filter wheels with 8 excitation and 8 emission filter positions
LUMIstar Omega: - Emission filter wheel with 8 filter positions
Gain control - Software selectable gain
- Automatic gain adjustment
Plate carrier - Auto lock microplate carrier
- All microplate formats up to 1536-well -
Microplates should fulfill the SBS specification and non-SBS formats
should fit: (l
x
w
x
h) (mm) max: 128
x
86
x
22; min. length 124
Reagent Injection - Up to two built-in reagent injectors
- Individual injection volumes for each well
- Injection volumes definable down to 3µl
- Up to 4 independent injection actions per well
- Variable injection speed (100 µl/s to 420 µl/s)
Shaking - Linear, orbital and double-orbital shaking
- User-definable time and speed
Incubation - Incubation range from ambient +4°C to 45°C, in 0.1°C steps
- Extended incubation up to 65°C (optional)
- Temperature monitoring (without incubation)
- Temperature stability 0.2°C
Fluorescence intensity - Limit of detection 0.2 fmol / well fluorescein
- Spectral range (ex. and em.): 240 ... 740 nm (900 nm opt.)
- Dynamic range 8 decades
Time-resolved fluorescence - Limit of detection 30 amol / well europium
- Advanced TRF/TR-FRET: 3 amol / well europium
- Spectral range (ex. and em.): 240 ... 740 nm (900 nm opt.)
- Dynamic range: 6 decades