14
Aesculap
®
High frequency surgical device GN060
5.5 Wipe disinfection for electrical devices without sterilization
RT: Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
5.6 Inspection, maintenance and checks
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Inspect all cables, and in particular electrode cables, for any damage to
their insulation.
Set aside the product if it is damaged.
Perform a safety inspection annually, see Maintenance.
5.7 Storage and transport
Transport the product only in its original box.
Observe storage and transport conditions, see Ambient conditions.
6. Maintenance
The HF device GN060 does not require maintenance.
Note
Maintenance must only be carried out by authorized personnel. If neces-
sary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
6.1 Safety inspection
A safety inspection must be carried out annually.
The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection proto-
col.
Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are
authorized to do so.
If there are significant deviations from the inspection protocol values,
or if the prescribed maximum values are exceeded: Send in device.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Phase Step T
[°C/°F]
t
[min]
Conc.
[%]
Water quality Chemical
IWipe disinfectionRT 1 - - Meliseptol HBV wipes 50 % Propan-1-ol
CAUTION
Damage to the product caused by switch-on to
early after storage/transport at temperature below
+10 °C (50° F)!
Allow the HF device to acclimate at room tem-
perature for about 3 hours.