High frequency surgical device GN060
7. Troubleshooting list
Error codes
In the event of an error. one of the error codes detailed in the following table appears in the display 8.
Error code Detection Cause Remedy
F1 Multiple keys pressed Several buttons activated at the
same time or object that presses on
several buttons
Only press one button or, if neces-
sary, remove object from the key-
board. If an error is still displayed,
there is probably a defect present
in the keyboard. Contact Aesculap
Technical Service.
F2 Membrane key held for longer than
15 s
A key has been activated for more
than 15 s or an object is pressing on
the keyboard
Activate the keys for a shorter
period of time or, if necessary,
remove the object from the key-
board. If an error is still displayed,
there is probably a defect present
in the keyboard. Contact Aesculap
Technical Service.
F3 Foot switch activated for longer
than 30 s
– Release foot switch and activate it
again. If the error message contin-
ues to be displayed even if the foot
switch is not activated, it is possi-
ble that this or the lead cable of the
foot switch defective. Contact
Aesculap Technical Service.
F4 EEPROM writing error It was not possible to write
EEPROM
Switch off the HF device and
switch it on again If the error mes-
sage continues to be displayed,
there is an error in the EEPROM.
Contact Aesculap Technical Ser-
vice.
F5 EEPROM reading error It was not possible to read EEPROM see F4
E1-E6 Error in electronics – Switch off the HF device and
switch it on again If the error mes-
sage continues to be displayed,
there is a device error. Contact
Aesculap Technical Service.
EF Microprocessor error –