68
Chapter 11
INSTRUCTIONS FOR USE ACCESSORY
Note: A maximum of three plugs can be stacked upon each other in socket P2.
Drop Sensor SP (8713175)
The drop sensor provides an additional safety function and is therefore
particularly recommended in connection with low delivery rates (10 ml/h).
The connection of the Drop Sensor SP on the pump is located at the rear of the
device, in the lower left corner. At the time of delivery the port of the drop
sensor is protected by a cover. Use a screw driver to break off the cover for
further disposal.
Use holder on PoleClamp, in order to park the drop sensor.
Short Stand SP (8713135)
Use the Short Stand SP to attach an infusion container to the pump.
1.)Push the PoleClamp on the pump.
2.)Plug the short stand into the aperture on the PoleClamp; make sure
that it audibly locks in.
3.)To remove the short stand: Press the white button at the lower end of
the PoleClamp and remove the short stand.
Note: Use only one infusion bag with max. 1000ml on the short stand.
Note: Insure pump is properly secured before attaching fluid bag to short stand
and connecting to patient to assure pump cannot fall and harm patient.
Battery-Pack SP (NiMH) (8713180)
Battery-Pack SP (NiMH) incl. Pin (8713180A)
For further information on the Battery-Pack SP (NiMH) see “Battery Operation”.
Battery-Pack SP (LiIon) incl. Pin and WiFi (8713182A)
For further information see Instructions for Use of "Battery Pack SP with WiFi".
Interface Lead CAN SP (8713230)
Interface Lead CAN SP is needed in order to set up a connection between the
SpaceStation/pump and the computer outlet (for service requirements).
1.) Push plug into socket F3 on the SpaceStation or P2 on the pump and
connect with the CAN/USB converter.