69
2.) Connect CAN/USB converter to computer outlet as described in the
Instructions for Use manual.
Caution: The Interface Lead CAN SP is only to be used by the service
department; never use while patient is connected.
Note: A maximum of three plugs can be stacked upon each other in socket P2.
Interface Lead RS232 SP (8713234)
Interface Lead RS232 SP is needed in order to set up a connection between the
Space pump and the computer outlet (for service requirements).
1.)Push plug into socket P2 on the pump and connect with the Interface
Lead RS232 SP.
2.)Connect Interface Lead RS232 SP to computer outlet as described in the
Instructions for Use manual.
Caution: The Interface Lead RS232 SP is only to be used by the service
department; never use while patient is connected.
Note: A maximum of three plugs can be stacked upon each other in socket P2.
Connection Lead SP (12 V) (8713231)
Install the Connection Lead SP (12 V) in the following way:
1.) Connect plug to socket P2 on back of pump or F3 on SpaceStation
respectively.
2.) Put the connection lead into the car socket.
3.) If necessary, remove red adaptor of motor vehicle connector by slightly
turning and simultanously pulling.
The green LED of the electronic box shows the operating voltage. The mains
connector can easily be replaced by another plug if required.
Caution: Do not connect the pump to a patient during external car battery
charging!
Note: A maximum of three plugs can be stacked upon each other in socket P2.
Connection Lead for Staff Call SP (8713232)
To connect the InfusomatĀ® Space P to staff call, use the Connection Lead for
Staff Call SP. The staff call needs to comply with the requirements of VDE 0834
(consider country specific regulations).
Chapter 11
INSTRUCTIONS FOR USE ACCESSORY