5.5 Using an External Battery
The iSleep 20+ can be powered from:
The Breas external EB 2 Battery Pack.
A 12 V external power source. The iSleep DC converter 12-24 V needs to be
used.
Battery Operation Time
The operation time is dependent on the battery condition, its capacity, the ambi-
ent air temperature and the iSleep 20+ pressure setting.
The operation time is based on the following settings:
1
Connect the external DC cable to
the iSleep 20+. Make sure it is fitted
correctly.
2
Connect the other end of the cable to the battery source.
Always make a function check to test the battery condition before operat-
ing the iSleep 20+ from an external battery source.
The battery must be disconnected when the iSleep 20+ is switched off,
otherwise the battery will be discharged.
SETTING VALUE
Humidifier setting 5 or Off
CPAP pressure 10 cmH
2
O
Ambient temperature 20°C (68°F)
Altitude Sea level
Battery status New battery fully charged
BATTERY TYPE HUMIDIFIER OPERATION TIME
EB 2 (24 V, 7.2 Ah) Off 13 h