38 Cleaning the iSleep 20+ and Replacing the Filters
iSleep 20+ users manual
Doc. 004270 En-Eu T-3b
Always clean the parts or use a new set when used by a new patient.
All parts that come into contact with the respiration gas must be cleaned as fol-
lows:
1
Place the dismantled parts in hot water containing washing-up liquid.
2
Remove fouling with a brush.
3
Rinse parts thoroughly under running hot water.
4
Shake water out of all parts.
5
Dry the parts completely.
6
Store in dust-free location.
The patient circuit shall be replaced in intervals according to the care provider’s
instructions. Check the patient circuit for damage regularly. In case of damage,
replace the circuit.
9.2 Cleaning and Replacing the Patient Air Filter
Washable Filter (grey colour, mandatory)
Replace the washable filter at least once a year. Wash the filter at least once a
week.
1
Wash the filter using warm water and a mild detergent.
2
Rinse thoroughly.
3
Dry the filter by squeezing it out in a towel. Do not wring the filter.
Disposable Filter (white colour, optional)
Replace the filter at least every 4th week, or more often when used in high pol-
lution or pollen-rich environments.
The patient air filter is located on the
side of the iSleep 20+. There are two
types of filters:
Do not wash or reuse the disposable filter.