30 Preparing the iSleep 20+ for Use
iSleep 20+ users manual
Doc. 004270 En-Eu T-3b
6.4 Connecting the Patient Circuit
Mask Leakage
The leakage from the mask should be at least 12 l/min at 4 cmH
2
O, to prevent
re-breathing of exhaled air. The recommended mask leakage is 20-40 l/min at
10 cmH
2
O pressure.
This leakage may be achieved by:
• small holes in the mask
• an adjoining ventilation valve
3
Connect the power cord to the
mains supply.
Read the chapter “Usage of Patient Circuit” on page 10 carefully to make
sure all conditions are fulfilled and considered.
1
Connect the patient tube to the
air outlet.
2
Connect the other end of the patient tube to the mask and the ventilation
valve, if applicable.
In order to receive a suitable leakage the Breas iMask™ Nasal CPAP Mask is
recommended.