41Technical Specifications
iSleep 20+ users manual
Doc. 004270 En-Eu T-3b
11 Technical Specifications
11.1 Data
The iSleep 20+ and it’s packaging do not contain any natural rubber latex.
SETTING/VALUE RANGE/SPECIFICATION RESOLUTION
CPAP 4 to 20 cmH
2
O0.5 cmH
2
O
Ramp time 5 to 60 min 5 min
Ramp start pressure 4 cmH
2
O to CPAP 0.5 cmH
2
O
Humidifier 1 to 9, 10 to 30 mgH
2
O/l,
<100% RH. Heat-up time
from 23°C: 1 hour. Max gas
temperature at patient port:
41°C.
1
Maximum flow >125 l/min
Maximum limited
pressure during sin-
gle fault condition
30 cmH
2
O
Max flow in CPAP
mode
6.7 cmH
2
O: 108 l/min
13.3 cmH
2
O: 132 l/min
20 cmH
2
O: 132 l/min
Breathing resist-
ance under single-
fault
0.9 cmH
2
O at 30 l/min
3.2 cmH
2
O at 60 l/min
Digital output 0 to 5 V
Sound level at 10
cmH
2
O
Less than 30 dB(A). Measured
at 1 m.
INDICATIONS SPECIFICATION TOLERANCE
Pressure 0 to 20 cmH
2
O
Power fail indica-
tion
Message shown in the display.