9Safety Information
iSleep 20+ users manual
Doc. 004270 En-Eu T-3b
2.3 Environmental Conditions
Do not use the iSleep 20+ in any toxic environment.
Do not use the iSleep 20+ in environments where there are explosive gases
or other flammable anesthetic agents present.
The performance of the iSleep 20+ may deteriorate at:
ambient temperatures below 5°C (41°F) and above 40°C (104°F).
ambient relative humidity below 10% RH (relative humidity) and above
95% RH.
atmosphere pressure below 700 mbar and above 1060 mbar.
Do not use the iSleep 20+ while positioned in a warm place, such as direct
sunlight.
The device complies with the EMC requirements of standards. Measures
should include but not be limited to:
normal precautions with regard to relative humidity and conductive char-
acteristics of clothing in order to minimize the build-up of electrostatic
charges.
avoiding use of radio emitting devices closer than 1 m to the iSleep 20+.
Radio emitting devices are for example cellular or cordless telephones,
microwave ovens and high-frequency surgery apparatus.
The iSleep 20+, any accessories and all replaced parts must be disposed of
in accordance with the local environmental regulations regarding the dis-
posal of used equipment and waste.
The performance of the iSleep 20+ and treatment of the patient may dete-
riorate if the operation conditions in “Technical Specifications” on page 41
are not fulfilled. Do not use the iSleep 20+ immediately after storage or
transport outside the recommended operating conditions.