EasyManua.ls Logo

Bristol 871A - Page 51

Bristol 871A
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 871A User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 51
SENSe Subsystem
The SENSe subsystem { XE “Commands:SENSe” } allows changes to how the instrument
makes measurements.
SENSe Subsystem Functions
:AVERage:STATe?
Description
Queries the state of the averaging status.
Query Response
{ OFF | ON }
Examples
:SENS:AVER:STAT?
:AVERage:STATe { OFF | ON}
Description
Sets the state of the averaging status.
*RST value
OFF
Examples
:SENS:AVER:STAT OFF
:AVERage:COUNt?
Description
Queries the number of readings being averaged for wavelength and
power values.
Query Response
{ OFF | 2 | 3 | 4 | 128 }
Examples
:SENS:AVER COUN?
:AVERage:COUNt { 2 | 3 | 4 | 128 }
Description
Sets the number of readings being averaged for wavelength and
power values.
*RST value
2
Examples
:SENS:AVER:COUN 8
:AVERage:DATA? { :POWer | :FREQuency | :WAVelength | :WNUMber }
Description
Returns averaged wavelength or power data for the last N number of
measurements. The value of N is set by :AVERage: COUNt command.
If averaging is not turned on then the most recent data is returned.
Examples
:SENS:AVER:DATA? POW
:SENS:AVER:DATA? WAV
:CALIbrate
Description
Initiates a calibration of the instrument.
Examples
:SENS:CALI