Model 871A User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 52
:CALIbrate:METHod { OFF | TIME | TEMPerature }
Description
Sets the method of automatic calibration for the instrument.
*RST value
TEMPerature
Examples
:SENS:CALI:METH TIME
:CALIbrate:METHod?
Description
Queries the method of calibration for the instrument.
Query Response
{ OFF | TIME | TEMPerature }
Examples
:SENS:CALI:METH?
:CALIbrate:TEMPerature { 1 | 2 | 3 | ... | 49 | 50 }
Description
Sets the minimum change in internal temperature that will trigger a
calibration of the instrument. The change is specified in 1/10
th
°C
increments
*RST value
5
Examples
:SENS:CALI:TEMP 25
:CALIbrate:TEMPerature?
Description
Queries the minimum change in internal temperature that will initiate a
calibration of the instrument. The change is specified in 1/10
th
°C
increments.
Query Response
{ 1 | 2 | 3 | ... | 49 | 50 }
Examples
:SENS:CALI:TEMP?
:CALIbrate:TIMer { 5 | 6 | 7 | ... | 1439 | 1440 }
Description
Sets the time between automatic calibrations of the instrument. The
time is specified in minutes.
*RST value
30
Examples
:SENS:CALI:TIM 100