Model 871A User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 56
:PID:VOLTage:MINimum?
Description
Queries the minimum voltage of the PID calculation, expressed in
volts.
Query Response
{ -5.0…0.0 }
Examples
:SENS:PID:MIN?
:PID:VOLTage:MAXimum { 0.01…5.0 }
Description
Sets the maximum output voltage for the PID calculation, expressed in
volts. The instrument automatically limits this voltage to a maximum of
5 volts.
*RST value
5.0
Examples
:SENS:PID:VOLT:MAX 2.5
:PID:VOLTage:MAXimum?
Description
Queries the maximum voltage of the PID calculation, expressed in
volts.
Query Response
{ 0.01…5.0 }
Examples
:SENS:PID:VOLT:MAX?
:PID:VOLTage:OFFset { -5.0…5.0 }
Description
Sets the voltage offset (centering position) for the PID calculation,
expressed in volts. The instrument automatically limits this voltage to
±5 volts.
*RST value
0.0
Examples
:SENSE:PID:VOLT:OFF 2.0
:PID:VOLTage:OFFset?
Description
Queries the voltage offset (centering position) for the PID calculation,
expressed in volts.
Query Response
{ -5.0…5.0 }
Examples
:SENS:PID:VOLT:OFF?
:PID:VOLTage:SCALe { -500.0…500.0 }
Description
Sets the voltage/gain scale for the PID calculation, expressed in volts
per nanometer.
*RST value
1.0
Examples
:SENSE:PID:VOLT:SCAL 20.0