Model 871A User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 57
:PID:VOLTage:SCALe?
Description
Sets the voltage/gain scale for the PID calculation, expressed in volts
per nanometer.
Query Response
{ -500.0…500.0 }
Examples
:SENS:PID:VOLT:SCAL?
:EXPOsure:AUTO { ON | OFF }
Description
Sets the state of the Auto Exposure function.
*RST value
OFF
Examples
:SENS:EXPO:AUTO OFF
:EXPOsure:AUTO?
Description
Queries the state of the Auto Exposure function.
Query Response
{ON | OFF}
Examples
:SENS:EXPO:AUTO?
STATus Subsystem
The STATus subsystem { XE “Commands:STATus” } is used to retrieve status information
from the instrument and to mask or screen what status can be retrieved.
Status registers give information about the instrument and the latest measurement. Status
enable registers allow some status information to be ignored. Status information disabled
though the status enable registers will not show up in the status registers.
STATus Subsystem Functions
:QUEStionable:CONDition?
Description
Queries the SCPI Questionable Status Register which contains bits
that indicate that one or more measurement types are of questionable
accuracy. The bits in the register are described in the table below.
Query Response
An integer which is the sum of the bit values for all bits in the register
that are set.
Examples
:STAT:QUES:COND?