Workaround
Swap logically both ethernet devices of the O2. That means, connect the
spect to ether-
net controller
ec0 on the motherboard of the O2. This can easy be carried out by modify-
ing the file
netif.options.
Type on O2 as superuser:
vi /etc/config/netif.options
This file contains the 2 parameter entries of if1name=ec0 and if2name=ec1. Swap the
values of both parameters to:
if1name=ec1
if2name=ec0
Type:
:wq
init5
When the O2 is shut down swap also the connections, first connection between spect
and motherboard and the second between the general net and the additional interface and
reboot the O2 and than the
spect.