82
ZTKS0243 / Z31993 / 03
Technical Data MS 400‘89 Ascend DNP
A.5.2 Technical Data Magnet
Operating modes of the Magnet System:
Driven Mode
In the driven mode the current lead is mounted and the electricity is flowing through the
power supply. The coils of the magnet can be energized and de-energized.
Persistent Mode
In the persistent mode the electricity is flowing exclusively within the magnet. The circuit
has no connection to the outside. The magnetic field cannot be switched off.
Technical Data Magnet Value Unit
Proton frequency 400 MHz
Central field 9.39 T
Coil inductance 59.5 H
Magnetic energy 527.9 kJ
Maximum drift rate 0.01 ppm/h
4Hz/h
Table A.9: Specification of the Magnet