Troubleshooting
52
Figure 45. Channel A Connection Error
Figure 46. Channel B Connection Error
Channel errors that may occur and the troubleshooting steps to take are provided in the
table below.
connector.
E309
E310
E312
E322
the Infusion
Catheter.
Infusion Catheter
Connector is wet at
the connection site.
1. Disconnect the Infusion Catheter.
2. Inspect for fluid or damage to the
pins.
3. Dry the Infusion Catheter
connector, if necessary.
4. Reconnect the Infusion Catheter
and ensure the connection is
secure.
5. Resume ultrasound.
EKOS Help Line.
E303
E316
E317
E318
E319
E323
E324
E325
E326
E327
E328
E329
E330
representative or call
+1.425.415.3100.