EKOS™ Control System 4.0
Instructions for Use
Caution: Federal law restricts this device to use by or on the order of a physician.
Part number: 8490099 Rev J

Questions and Answers

  • A
    Antonio KramerSep 12, 2025
    What does it mean if Infusion Catheter and Ultrasonic Core are not compatible in BTG Control Systems?
    • J
      Jonathan RossSep 12, 2025
      If the BTG Control Systems indicates that the Infusion Catheter and Ultrasonic Core are not compatible, it means there's a mismatch in either catheter zone length, ultrasound zone length, or working length. Notify the physician about this issue.
  • N
    Nicole MoranSep 12, 2025
    How to resolve incompatibility of CIC error on BTG EKOS Control Systems?
    • P
      Patricia MorrisonSep 13, 2025
      If the BTG Control Systems displays a message about incompatibility of CIC, it means the Connector Interface Cable is not compatible with the Infusion Catheter, Ultrasonic Core, or both. Replace the Connector Interface Cable with one that is compatible with both the Infusion Catheter and Ultrasonic Core.
  • J
    James JohnstonSep 15, 2025
    Why does BTG EKOS say to replace the Connector Interface Cable?
    • H
      harrisbrandonSep 15, 2025
      If you are prompted to replace the Connector Interface Cable on your BTG Control Systems, this indicates that the cable is defective. Replace the Connector Interface Cable.

Summary

General Information

EKOS Control System 4.0

Information on the Control Unit, cables, and its functions for ultrasound therapy.

System Components and Interface

Safety Information

Warnings and Precautions

Crucial safety statements for operating the Control Unit to prevent injury or damage.

Power Management

AC Power and Battery Operation

Details AC power use, battery charging, and status indicators.

Battery Status and Alerts

Explains battery gauge, low battery, critically low battery, and error notifications.

Ultrasound Therapy Operation

Setup and Starting Therapy

Guides through preparing the system and initiating ultrasound therapy.

Performance Monitoring

Troubleshooting

System and Channel Errors

Identifies system and channel errors, their causes, and resolution steps.

Specifications

BTG EKOS Specifications

General IconGeneral
ManufacturerBTG
ModelEKOS
Input Voltage24 VDC
Power Consumption10 W
Communication ProtocolModbus RTU
Protection ClassIP20
CategoryControl Systems
Operating Temperature0 to 50°C