3 Technical data
12
NIRMaster Operation Manual, Version B
3 Technical data
This section introduces the reader to the instrument specifications. It contains the technical data and
materials used.
Information on type plate:
a
b
c
d
e
f
a
Instrument type code
Example:
NIRM54P = NIRMaster + 54 (IP class) + P (Pro version, steel)
b
Serial Number
c
Supply voltage range / type
d
Frequency of supply voltage
e
Nominal power rating
f
Year of manufacture
Technical data NIRMaster
Electrical connection data
IP 54 IP 54 Pro IP 65 Pro
Voltage 100 – 240 VAC ±10% 100 – 240 VAC ±10% 100 – 240 VAC ±10%
Frequency 50/60 Hz 50/60 Hz 50/60 Hz
Max. power consumption 150 W 150 W 150 W
IP class 54 54 65
Installation category II II II
Internal fuse 4 A / 240 V
(2× resettable automatic fuses, all poles)
Environmental conditions (indoor use only)
Pollution degree
2
Temperature
5 – 35 °C (25 ±5 °C recommended)
Altitude (above sea level)
up to 2000 m
Humidity (curve parameter)
Maximum relative humidity < 80% up to < 31 °C, then decreasing linearly to
67% relative humidity at 35 °C