8 Troubleshooting
59
NIRMaster Operation Manual, Version B
8.3 Reference measurement fails
When the deviation between the stored reference spectrum is out of limit, the reference measurement
will fail. A pop-up window will show both curves and give details of the measurement result.
Possible reasons for a reference measurement to fail
Optical surfaces (e.g. External reference) dirty or scratched
Environmental conditions out of operational range (e.g. condensating humidity at optical surfaces)
Defective instrument (e.g. electronic damage)
After lamp-module exchange
Remedy
Clean the External reference and the NIRMaster output window
Place the External reference onto the output window and press the ‘Rerun’ button to repeat the
measurement
If measurement still fails, a new reference spectrum can be stored by pressing the ‘Checkmark’
button. A login window will pop-up. An administrator login is required to store the new data set!