6 Operation
30
NIRMaster Operation Manual, Version B
6.3.2 Frequency of reference measurement
An additional External reference measurement is mandatory:
after application change
after lamp exchange and lamp failure
after service (when top cover has been reinstalled and locked!)
when changing the add-on or measurement setup
at every system start / restart
6.4 Operator software – the user interface
Every NIRMaster comes installed with two different user interfaces of the Operator software.
Operator NIRMaster
Operator NIRMaster is optimized for touch-screen usability in production environment (large buttons,
extra large character height). Reduced to the basic functionality of retrieving a result in the least neces-
sary amount of steps the program is wizard guided. The displayed set of applications depends on the
user rights and can be configured to minimize the possibility of handling errors.
Operator
Operator (extended functionality, but without wizard) is optimized for mouse and keyboard use.
Offering mostly the same functionality as the NIRMaster Operator, the additional software buttons
(compared to the Operator NIRMaster view) listed below are accessible.
Buttons main window
Button Function key Functionality
n.a.
‘Change user’ log off and in again as a different user
F 4
‘Service tool’ enter service tools view
F 5
‘Logger’ open logfile viewer
n.a.
‘Select application’ open list of applications
n.a.
‘LIMS import’ import LIMS files (system must be connected to a LIMS)