BÜCHI Labortechnik AG Product description | 3
Operation Manual Vacuum Pump V-100 15/52
3.4 Technical data
3.4.1 Vacuum Pump V-100
Dimensions (W x H x D) 180 x 275 x 210 mm
Weight 5.6 kg
Power supply voltage 100 – 240 V, 50/60 Hz
Power consumption 150 W
Power consumption in ECO mode 70 W
Terminal vacuum (absolute) 10 mbar (± 2 mbar)
Pump capacity 1.5 m
3
/h
Vacuum connection GL-14
Motor Brushless DC motor
Rotation speed 1280 rpm max.
Rotation speed in ECO mode 70% of max. capacity
Noise level 32 to 57 dBA
57 dBA: with 100% load
Overvoltage category II
Pollution degree 2
Enclosure rating IP20
Certificates CE
3.4.2 Ambient conditions
Max. altitude above sea level 2000 m
Ambient temperature 5 - 40 °C
Maximum relative humidity 80 % for temperatures up to 31 °C
decreasing linearly to 50 % at 40 °C
The Vacuum Pump V-100 may only be used in indoor areas.
3.4.3 Materials
Component Material
Pump heads PPS
Drive unit body Aluminium
Housing PBT
Membranes PTFE/rubber
Valve plate PEEK
Valve head PEEK
Pump-head connection piece FEP
Vacuum tubing Silicone/natural rubber
Non-return valve O-rings FFKM

Table of Contents