BÜCHI Labortechnik AG Installation | 5
Operation Manual Vacuum Pump V-100 19/52
5.2 Installation site
u Before setting up the V-100 and putting it into service, remove the transit fixings
(red) on the rear of the device (see Chapter4.3 "Removing transit fixings",
page17).
IMPORTANT
Risk of property damage from falling equipment (e.g. in an earthquake)
u Secure the Vacuum Pump against falling by means of the lashing eye on the rear
of the device (see Chapter3.2.2 "Rear view", page11).
The installation site must meet the following requirements:
Firm and level surface: at least 200 mm x 300 mm (W x D)
Height: at least 300 mm
Set up inside a fume hood or
feed the pump outlet into the fume hood
Sufficient air supply to the pump cooler
The V-100 is intended for use in laboratory environments (see Chapter3.4.2 "Ambi-
ent conditions", page15).
5.3 Connecting laboratory equipment