BÜCHI Labortechnik AG Installation | 5
Operation Manual Vacuum Pump V-100 23/52
5.5 Connecting a Woulff bottle
1
5
2
4
6
8
9
10
3
7
Fig.10: Connecting a Woulff bottle to the V-100
1 GL14 cap nut 6 Pump intake
2 Woulff bottle outlet 7 Connection piece
3 Top outlet for connecting Interface
I-100
8 Screw
4 Woulff bottle inlet 9 Mounting rail for holder
5 Woulff bottle glass component 10 Woulff bottle holder
The Woulff bottle is connected to the pump intake on the rear of the V-100.
Tools required: Torx key (fitted in pump top cover)
u Fix the mounting rail (9) for the holder below the pump intake (6) using the screw
(8).
u Fit the holder (10) onto the mounting rail.
u Unscrew the GL14 screw-cap from the pump intake (6).
u Fit the connection piece (7) from the outlet of the Woulff bottle onto the pump in-
take. Screw the GL14 cap nut (1) on the end of the tubing onto the pump intake
thread.
u Press the Woulff bottle into the holder.
Connecting an Interface I-100 via a Woulff bottle
u If necessary connect the Woulff bottle to the Interface I-100 as follows: Fit the con-
nection piece to the top outlet (3) of the Woulff bottle.
u For details of connecting the connection piece to the interface unit and the setup,
also refer to the operating instructions for the I-100.