5 | Installation BÜCHI Labortechnik AG
20/52 Operation Manual Vacuum Pump V-100
The V-100 is connected by a vacuum tubing to the laboratory apparatus to be evacu-
ated. The vacuum tubing leads via a GL14 connection from the pump intake (2) to the
outlet of the laboratory device concerned.
Precondition:
R Transit fixings have been removed (see Chapter4.3 "Removing transit fixings",
page17).
u Fit tubing to pump intake (2) using the GL14 cap nut.
If vapors are to be expected during evacuation, we recommend the following:
u Connect a Woulff bottle upstream of the pump intake (see Chapter5.5 "Connect-
ing a Woulff bottle", page23).
u Connect the tubing to the Woulff bottle intake using a GL14 cap nut.
For details of connecting the V-100 to a rotary evaporator (e.g. Rotavapor® R-100)
see the relevant section of the device's operating instructions.
For more information on the use of the V-100, see Chapter3.2.4 "Typical application",
page13.