6.3. Fluid data
Chlorine sensor
568524
1)
Chlorine sensor
568523
2)
"Trace" chlorine sensor
565164
Type of medium • swimming pool water, drinking
water, service water, process
water
• free of any surfactants
• with constant pH value.
• swimming pool water, drinking
water, sea water
• surfactants are partially
tolerated.
water with drinking water quality
Medium temperature 0...+45 °C 0...+45 °C 0...+40 °C
Maximum medium pressure 1 bar, no pressure drops are
allowed
3 bar, no pressure drops are
allowed
0,5 bar, no pressure drops are
allowed
Medium pH range 6...8 pH 4...9 pH 6,5...9 pH
Flow rate approximately 15...30 l/h approximately 15...30 l/h approximately 15...30 l/h
Temperature compensation automatic, with integrated tem-
perature sensor
automatic, with integrated tem-
perature sensor
automatic, with integrated tem-
perature sensor
Measuring range at nominal
slope
0,01...20 ppm 0,01...20 ppm 0,005...2 ppm
Response time (t90) 30 s 120 s 120 s
Run-in time after start-up
maintenance
1 h 2 h 2 h
Standing time of the electrolyte up to 6 months
(depending on water quality)
up to 12 months
(depending on water quality)
up to 6 months
(depending on water quality)
1) or other chlorine sensor for operation at constant pH
2) or other chlorine sensor having a reduced pH-dependency