EasyManua.ls Logo

Burkert 8232 - Installation of the Analytical Measurement Chamber Type 8200; Installation of the Sensor into the Analytical Measurement Chamber

Burkert 8232
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.4. Installation of the analytical
measurement chamber
type 8200
DANGER
Risk of damaging the sensor due to high temperatures or
pressure.
Respect the sensor pressure and temperature ranges.
Close the water inlet ball valve of the analytical measure-
ment chamber type 8200.
Avoid installations that allow air bubbles to enter the
measuring water, because air bubbles must be avoided
on the membrane.
Mount the analytical measurement chamber type 8200 with
two screws (recommendation: M4 x 60 mm pan head screw or
hexagon socket head screw) onto a suitable mounting panel.
Connect the water inlet of the analytical measurement chamber
with a 6...8 mm hose to the sample water source.
Connect the water outlet of the analytical measurement chamber
with a 6...8 mm hose to the drain for example.
For more information, refer to the datasheet of the analytical meas-
urement chamber type 8200, available at: www.burkert.com.
7.5. Installation of the sensor into
the analytical measurement
chamber
Depressurise the system.
Unscrew the nut of the analytical measurement chamber
type 8200.
Make sure the snap ring, the slide ring and the seal are in their
correct position on the sensor.
Insert the prepared sensor into the analytical measurement
chamber type 8200.
Insert the nut on the sensor and screw it tightly on the analytical
measurement chamber.
Carefully open the inlet ball valve so that the float just touches
the upper end of the flow indicator.
Wire the sensor according to chap. 7.6.
18
Installation and wiring
Type 8232
english

Table of Contents

Related product manuals